Protein Data Bank exchange dictionary (pdbx) version 1.0521
_em_detector.detective_quantum_efficiency
Name:'_em_detector.detective_quantum_efficiency'
Definition:
The detective_quantum_efficiency (DQE)is defined as the
square of the signal-to-noise ratio in the recording device
divided by the square of the signal-to-ratio in the electron beam:
(SIGNAL/NOISE)2 recording device
DQE = -------------------------------
(SIGNAL/NOISE)2 electron beam
A DQE value of 1 indicates a perfect recorder. "DQE = 0.25" menas
that the signal-to-noise ratio is reduced by half in the
recording step.
(0.5)**2
DQE = --------- = 0.25.
(1.0)**2
Example:| 0.25 |
Type: float
Mandatory item: no
Category: em_detector


![[CIF home page] [CIF logo]](https://www.iucr.org/__data/assets/image/0015/131037/CIF_white.png)